Ortivus AB 410000409 User Manual

Page of 47
72
10 Appendix
10.4.3 Batteries
Problem
Cause
Action
The 300c batteries 
can not charge.
The 300c does not 
receive external 
power.
Make sure that the two indicators for external power 
supply connected and battery charging are lit, see 
“Units in the system” on page 5.
300c is too cold or hot 
to charge.
Warm or cool 300c in order to charge 300c battery.
The operating time of 
the 300c batteries is 
low.
Batteries are getting 
old.
Leave to authorised personnel for replacing batteries.
The 300m batteries 
do not charge.
The 300m does not 
receive external 
power.
Make sure that the 300m receives external power.
Make sure that the 300m is correctly docked.
The operating time of 
the 300m is short.
Batteries are getting 
old.
Leave to authorised personnel for replacing batteries.