Appendix 10
10.4.4 ECG
Problem
Cause
Action
No ECG waveform is 
displayed in monitor 
view.
The 300c is not 
turned on or the 300m 
can not establish 
contact with the 300c
Check that the 300c is turned on and that contact is 
established.
ECG cable not 
connected.
Check that the ECG label is visible in monitor view, 
see “Monitor and workspace view” on page 12, 
otherwise ECG cable is not connected properly.
Disturbances and 
alarms on ECG in 
monitor view.
Poor ECG signal 
quality.
Check ECG signal quality in Electrode test view, see 
“Electrode test” on page 26. Check electrode contact, 
check if old electrodes, check that ECG cable is not 
damaged.
ECG alarms “ECG 
Learning” or 
“Selecting reference 
complex” wont go 
away.
Poor ECG signal 
quality.
Check ECG signal quality in Electrode test view. 
Check electrode contact; check if the electrodes are 
old, check that ECG cable is not damaged.
Pacemaker spikes 
are not shown in ECG 
waveforms.
Pacemaker option is 
not chosen.
Choose pacemaker option in ECG settings, see “ECG 
settings” on page 31.
No ECG complexes 
in ECG view.
Wrong ECG cable 
used.
For 3 lead cable no complexes can be calculated. 
Change ECG cable.
MIDA analysis not 
working.
Press Relearn MIDA in MIDA settings, see “MIDA 
settings” on page 32.
MIDA relearn failed.
Poor electrode 
contact.
Check ECG signal quality in Electrode test view. 
Check electrode contact, check if old electrodes, 
check that ECG cable is not damaged. Then try MIDA 
Relearn again.
Too few ECG 
complexes in ECG 
view.
Wrong ECG cable 
used.
For 5 lead cable all complexes cannot be calculated. 
Change ECG cable.