Ortivus AB 410000409 User Manual

Page of 47
74
10 Appendix
10.4.5 SpO
2
10.4.6 NIBP
10.4.7 Respiration
Problem
Cause
Action
No SpO
2
 waveform is 
visible in monitor 
view.
SpO
2
 probe not 
connected.
Check that the SpO
2
 label is visible in monitor view 
see “Monitor and workspace view” on page 12, 
otherwise SpO
2
 cable is not connected properly.
SpO
2
 probe not 
working.
Check that the SpO
2
 probe emits red light.
Disturbances and 
alarms on SpO
2
 in 
monitor view.
Bad SpO
2
 signal 
quality.
Disturbances can be caused by movement or if patient 
has poor circulation
 
in finger.
Problem
Cause
Action
No NIBP table is 
visible on monitor.
NIBP hose not 
properly connected.
Check that the hose is properly connected.
NIBP measurement is 
not started.
Start NIBP measurement, see “Start an NIBP 
measurement” on page 41.
Wrong NIBP cuff 
used.
Make sure that right size of NIBP cuff is used. See 
“Selection and placement of the cuff” on page 40.
NIBP cuff not properly 
placed.
Make sure that NIBP cuff is properly placed. See 
“Selection and placement of the cuff” on page 40.
Problem
Cause
Action
No respiration 
waveform is visible in 
monitor view.
Respiration 
measurement is not 
started.
Check that the Respiration label is visible in monitor 
view, otherwise start Respiration measurement in 
Respiration settings, see “Respiration settings” on 
page 47.
ECG cable not 
connected.
Check that the ECG label is visible in monitor view, 
otherwise ECG cable is not connected properly.
ECG electrodes poor 
contact.
Check the signal quality in Electrode test view (see 
“Electrode test” on page 26) of the ECG electrodes 
used for Respiration measurement, see “Electrodes 
for respiration measurement” on page 46. Adjust 
electrodes if necessary.
Disturbances and 
alarms on Respiration 
in monitor view.
Bad measurement 
environment.
The bioimpedance method is sensitive to movement 
by the patient, try using the 4-wire measurement, see 
“Electrodes for respiration measurement” on page 46.