Changing Survey Page Configuration 
 
Prior to first use in a care unit, the effective range and coverage 
area of each transmitter should be tested by conducting a paging 
survey to ensure that the transmitter and pagers are performing 
as needed. 
 
The programmer can be used to set the time between pages and 
the total number of pages to be sent for performing survey 
paging (the default in 100 pages, sent every 15 seconds). 
 
To set interval between pages and total number of pages: 
 
1. From SysMgr, tap Start button and tap Install Program. 
 
2. Enter password. Alarms screen will be displayed. 
 
3. Tap Monitor tab to display Monitor Configuration screen. 
 
{Insert monitor configuration screen (Figure 6-9)} 
 
Monitor Configuration Screen 
 
!
!
!
!
!
!
 
 
 
 
 
 
Monitor Type and Authorization fields are read-only values set at 
initial installation. You will not need to edit these values. 
 
4. To change survey page time intervals, tap Down Arrow in 
the sec (seconds) field. Tap to select desired time interval 
from pull-down menu.  
 
5. To change total number of survey pages, tap Down Arrow 
in the times field. Tap to select desired number of pages 
from pull-down menu. 
 
6. Program survey page configuration into transmitter. Turn 
transmitter on and aim infrared port on programmer directly 
at infrared port on front of transmitter and tap SEND button. 
Programmer will display a “Sending Program Data” status 
message.