Data Critical Corporation AVTS1010 User Manual

Page of 55
AlarmView™ System User’s Guide 
 
 
 
{insert part #} Page 37 
 
Transmitter Range Test (Survey Paging) 
 
Survey paging is used to determine the range and area of coverage of 
your system. An Transmitter Range Test (survey paging) will be 
performed at the time of initial system installation. Survey Paging should 
be conducted prior to first use of a transmitter added to a system, or 
when a transmitter has been moved to a new care unit. 
 
The transmitter is placed in Test Range Mode, and it will send frequent 
pages from the transmitter to the assigned pager(s). Frequency of sent 
pages and the total number of survey pages can be set according to the 
procedure in Section 4: Programming, “Changing Survey Page 
Configuration
” of this manual.  
 
As survey pages are sent, personnel with assigned pager(s) will walk 
around the coverage area of the care unit to ensure that all pages are 
received. 
 
!
!
!
 
 
 
Because transmitters and pagers in one system operate on a single, 
defined frequency, if you exchange pagers or transmitters from 
different systems, you must change transmitter or pager system 
configurations. Refer to Section 4: Programming, “Changing System 
Configuration”
 
To perform survey paging for a transmitter, transmitters should be turned 
on, but not communicating with monitors. You will need at least one 
transmitter and one pager. 
 
To conduct a transmitter range test (survey paging): 
 
1. Ensure that transmitter is connected to power and turned on, but 
not communicating with a monitor. 
 
2. On programmer, from SysMgr main screen, tap Service to enter 
service program. Controls tab will be displayed. 
 
3. To start survey paging, aim programmer infrared port directly at 
port on front of transmitter and tap Test Range button {insert test