Check monitor 
performance. 
Comm 
Failure/Shutting Down 
 
Transmitter has not 
been able to 
communicate with the 
monitor for about an 
hour and is shutting 
down. 
Communications 
between transmitter 
and monitor have 
failed. 
 
Monitor is turned off or 
operating on battery 
power and transmitter 
is still on. 
Check serial cable 
and cable connections 
between transmitter 
and monitor. 
 
Reconnect or replace 
cable as required. 
 
Connect monitor to 
AC power. 
 
Check monitor 
performance. 
 
The programmer may also display error messages related to its function. 
Below is a table of Programmer Error Messages, along with probable 
causes, and recommended corrective actions. 
 
 
Programmer Error Messages 
 
Message 
Possible Cause 
Corrective Action 
Failed to Connect 
 
Programmer was 
unable to retrieve or 
send information 
to/from transmitter. 
Programmer infrared 
port not aligned with 
transmitter infrared 
port. 
 
Transmitter turned off. 
 
Programmer batteries 
are low. 
 
Transmitter defective. 
 
Programmer 
defective. 
Re-align ports and try 
again. 
 
Turn on transmitter. 
 
Recharge programmer 
batteries. 
 
Contact Data Critical 
Corp. regarding repair 
and/or replacement of 
transmitter or 
programmer.