PillCam Capsule Endoscopy
228
Guidance and Manufacturer's Declarations
Recommended separation distances between portable and mobile RF communications 
equipment and the PillCam capsules
The PillCam capsules are intended for use in an electromagnetic environment in which radiated 
RF disturbances are controlled. The customer or the user of the PillCam capsules can help 
prevent electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the PillCam capsules as 
recommended below, according to the maximum output power of the communications 
equipment.
Rated maximum 
output power of 
transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2,5 GHz
d = 2.3 P
0.01
Not applicable
0.12
0.23
0.1
Not applicable
0.38
0.73
1
Not applicable
1.2
2.3
10
Not applicable
3.8
7.3
100
Not applicable
12
23
For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in meters (m) can be determined using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter in 
watts (W) according to the transmitter manufacturer.
Note 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range 
applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.