Biolog-id SSTABIOLOGID User Manual
Smart Storage Agitator User Manual
Version: 1.0
Validation date of the manual: 18/09/2018
24 of 27
6. First Level Maintenance
This chapter describes the first-level failures, which you can meet when using the SST-A.
Actions to be taken
Appearance of a red
indicator on the facades
of satellites
There is potentially no
traceability at the
location where the
indicator is red
1 Move the bags to a
functional location
2 Notify maintenance
reference personnel for
corrective maintenance
intervention.
corrective maintenance
intervention.
Appearance of red
indicator on the facade
of the calculator
Notify maintenance reference
personnel for corrective
maintenance intervention.
personnel for corrective
maintenance intervention.
Appearance of red and
green indicators on the
front of the calculator
Notify maintenance reference
personnel for corrective
maintenance intervention.
personnel for corrective
maintenance intervention.
When a red LED appears, try to detect the cause of the fault and eliminate it as soon as
possible.
Red LED on satellite
Possible causes
Action
Unable to write
Moving the bags to another location and try again
The drawer has been open for more than 4
minutes
Close the drawer
No communication with the calculator
Switch to maintenance mode and then restart the
calculator
calculator