5.  Cleaning Instructions 
This chapter explains the procedure for cleaning the SST-A. 
For proper operation, clean the SST-A at least once a month and more if necessary. 
Only  personnel  qualified  by  the  establishment  are  authorized  to  clean  the  SST-A.  Cleaning 
personnel should be aware of the operation of the SST-A and its documentation, especially 
the safety instructions. 
The cleaning must be done as follows; 
  Put the SST-A in maintenance mode. 
  Move the bags into another agitator. 
  Use a chemically compatible spray product with SST-A component materials, 
combining cleaning and disinfecting and scrubbing with a soft cloth 
 
Fig.  1.  Apply  the  disinfectant  detergent  spray  to  the  area  to  be  treated  or  to  a  nonwoven 
wipe. 
 
Fig.2. Distribute the product