electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C
customers or users must ensure that their devices are used in such an environment.
customers or users must ensure that their devices are used in such an environment.
Immunity test
Test level
According to CEI
60601
Level of
compliance
Electromagnetic environment - directives
Emissions -
Conducted RF
CEI 61000-4-6
Emissions -
Radiated RF
CEI 61000-4-3
3 Veff
150 kHz to 80 MHz
3 V
80 MHz to 2.5 GHz
3 V
3 V
Portable or mobile RF communication devices must not be
used any closer than the recommended separation
distance from any part of the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C, including the
cables, as calculated using the applicable equation for the
emitter frequency.
Recommended separation distance
d = 1.17 √
used any closer than the recommended separation
distance from any part of the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C, including the
cables, as calculated using the applicable equation for the
emitter frequency.
Recommended separation distance
d = 1.17 √
P
d = 1.17 √ P
80 MHz to 800 MHz
d = 2.34 √ P
800 MHz to 2.5 GHz
where P is the maximum characteristic power output of
the emitter in Watts (W) as provided by the manufacturer
of the emitter, and d is the recommended separation
distance in metres (m).
The strength of the fixed RF emitter fields, determined by
carrying out an electromagnetic survey on site ᵃ, must be
below the compliance level in each of the range of
frequencies. ᵇ
Interference can be caused by proximity to devices
marked with the following symbols:
NOTE 1 At 80 MHz and at 800 MHz, the highest range of frequencies is applicable.
NOTE 2 These directives might not be applicable to all situations. Electromagnetic propagation is affected by absorption and
reflections of structures, objects and persons.
ᵃ The strength of the fixed emitter fields such as radiotelephone base stations (cellular/cordless), mobile terrestrial radios,
amateur radios, AM and FM radio transmitters and TV transmitters cannot be calculated theoretically with any degree of
accuracy. To assess the electromagnetic environment due to fixed RF emitters, an on-site electromagnetic survey must be
carried out. If the strength of the field measured at the location in which the PRD-7130100C, PRD-7130110C, PRD-7130120C
or PRD-7130130C is used exceeds the applicable RF compliance level set out above, the PRD-7130100C, PRD-7130110C,
PRD-7130120C or PRD-7130130C must be monitored to check that it is operating normally. If the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C is observed to be performing abnormally, additional measures may be needed
such as re-orientating or repositioning it.
ᵇ The field strength over the range 150 kHz to 80 MHz must be lower than 3 V/m.
NOTE 2 These directives might not be applicable to all situations. Electromagnetic propagation is affected by absorption and
reflections of structures, objects and persons.
ᵃ The strength of the fixed emitter fields such as radiotelephone base stations (cellular/cordless), mobile terrestrial radios,
amateur radios, AM and FM radio transmitters and TV transmitters cannot be calculated theoretically with any degree of
accuracy. To assess the electromagnetic environment due to fixed RF emitters, an on-site electromagnetic survey must be
carried out. If the strength of the field measured at the location in which the PRD-7130100C, PRD-7130110C, PRD-7130120C
or PRD-7130130C is used exceeds the applicable RF compliance level set out above, the PRD-7130100C, PRD-7130110C,
PRD-7130120C or PRD-7130130C must be monitored to check that it is operating normally. If the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C is observed to be performing abnormally, additional measures may be needed
such as re-orientating or repositioning it.
ᵇ The field strength over the range 150 kHz to 80 MHz must be lower than 3 V/m.