electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C 
customers or users must ensure that their devices are used in such an environment. 
Immunity test 
Test level 
According to CEI 
60601 
Level of  
compliance 
Electromagnetic environment - directives 
 
 
 
 
 
 
 
 
Emissions - 
Conducted RF 
CEI 61000-4-6 
 
 
 
Emissions - 
Radiated RF 
CEI 61000-4-3 
 
 
 
 
 
 
 
 
3 Veff 
150 kHz to 80 MHz 
 
 
 
 
3 V 
80 MHz to 2.5 GHz 
 
 
 
 
 
 
 
 
3 V 
 
 
 
 
 
 
3 V 
Portable or mobile RF communication devices must not be 
used any closer than the recommended separation 
distance from any part of the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C, including the 
cables, as calculated using the applicable equation for the 
emitter frequency. 
 
Recommended separation distance 
d = 1.17 √
 
P 
 
d = 1.17 √ P
     80 MHz to 800 MHz
 
 
d = 2.34 √ P
   
800 MHz to 2.5 GHz
 
 
where is the maximum characteristic power output of 
the emitter in Watts (W) as provided by the manufacturer 
of the emitter, and is the recommended separation 
distance in metres (m). 
The strength of the fixed RF emitter fields, determined by 
carrying out an electromagnetic survey on site ᵃ, must be 
below the compliance level in each of the range of 
frequencies. ᵇ 
Interference can be caused by proximity to devices 
marked with the following symbols: 
 
NOTE 1 At 80 MHz and at 800 MHz, the highest range of frequencies is applicable. 
NOTE 2 These directives might not be applicable to all situations. Electromagnetic propagation is affected by absorption and 
reflections of structures, objects and persons. 
ᵃ The strength of the fixed emitter fields such as radiotelephone base stations (cellular/cordless), mobile terrestrial radios, 
amateur radios, AM and FM radio transmitters and TV transmitters cannot be calculated theoretically with any degree of 
accuracy. To assess the electromagnetic environment due to fixed RF emitters, an on-site electromagnetic survey must be 
carried out. If the strength of the field measured at the location in which the PRD-7130100C, PRD-7130110C, PRD-7130120C 
or PRD-7130130C is used exceeds the applicable RF compliance level set out above, the PRD-7130100C, PRD-7130110C, 
PRD-7130120C or PRD-7130130C must be monitored to check that it is operating normally. If the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C is observed to be performing abnormally, additional measures may be needed 
such as re-orientating or repositioning it. 
ᵇ The field strength over the range 150 kHz to 80 MHz must be lower than 3 V/m.