Recommended separation distances between portable and mobile RF communication devices and the PRD-7130100C, 
PRD-7130110C, PRD-7130120C or PRD-7130130C 
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in electromagnetic 
environments in which radiated RF emissions are controlled. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-
7130130C customers and users can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF communication devices (emitters) and the PRD-7130100C, PRD-7130110C, PRD-7130120C or PRD-
7130130C, as recommended below and in accordance with the maximum emitted power of the communication devices in 
question. 
Maximum assigned output power 
of the emitter 
Separation distance according to emitter frequency 
150 kHz to 80 MHz 
d = 1.17 √ P 
80 MHz to 800 MHz 
d = 1.17 √ P 
800 MHz to 2.5 GHz 
d = 2.34 √ P 
0.01 
 
 
 
0.1 
0.37m 
0.37m 
0.74m 
 
 
 
10 
 
 
 
100 
 
 
 
In the case of an emitter whose assigned maximum output power is not given above, the recommended separation 
distance in metres (m) can be estimated using the applicable equation for the emitter frequency, where is the maximum 
characteristic power output of the emitter in Watts (W) according to its manufacturer. 
NOTE 1 At 80 MHz and at 800 MHz, the separation distance for the highest range of frequencies is applicable. 
NOTE 2 These directives might not be applicable to all situations. Electromagnetic propagation is affected by absorption and 
reflections of structures, objects and persons. 
Directives and manufacturer declaration - electromagnetic immunity 
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in the 
electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C 
customers or users must ensure that their devices are used in such an environment. 
Immunity test 
Test level 
CEI 60601 
Level of  
compliance 
Electromagnetic environment - directives 
Electrostatic 
discharges 
(ESD) 
CEI 61000-4-2 
± 6 kV on contact 
± 8 kV into the air 
± 6 kV on contact 
± 8 kV into the air 
The floors must be made of wood, concrete 
or ceramic tiles. If the floors are covered with 
synthetic material, relative humidity must be 
at least 30%. 
Fast 
transient bursts 
 
CEI 61000-4-4 
± 2 kV for electrical 
supply lines 
± 1 kV for 
input/output lines 
± 2 kV for electrical 
supply lines 
± 1 kV for input/output 
lines 
The quality of the electrical power supply 
network must equate to that in a typical 
commercial or hospital environment. 
Transient 
overvoltages 
CEI 61000-4-5 
± 1 kV between 
phases 
± 2 kV between 
phase and earth 
± 1 kV between phases 
± 2 kV between phase 
and earth 
The quality of the electrical power supply 
network must equate to that in a typical 
commercial or hospital environment. 
Voltage dips, short 
outages and voltage 
fluctuations in the 
electrical power supply 
input lines 
 
CEI 61000-4-11 
<5% U
(>95% UV dip) for 
0.5 cycles 
<40 % U
(60 % UV dip) for 5 
cycles 
<70% U
(30% UV dip) for 25 
cycles 
<5% U
(>95% UV dip) for 5 
seconds 
<5% U
(>95% UV dip) for 0.5 
cycles 
<40 % U
(60 % UV dip) for 5 
cycles 
<70% U
(30% UV dip) for 25 
cycles 
<5% U
(>95% UV dip) for 5 
seconds 
The quality of the electrical power supply 
network must equate to that in a typical 
commercial or hospital environment. If a 
PRD-7130100C, PRD-7130110C, PRD-
7130120C or PRD-7130130C user requires 
continuous operation during electrical power 
supply network outages, we recommend 
connecting the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C 
to an uninterruptible power supply or 
battery. 
Magnetic field at the 
electrical network 
frequency (50/60 Hz) 
 
CEI 61000-4-8 
3 A/m 
3 A/m 
Magnetic fields at the electrical network 
frequency must be at levels that are 
characteristic of a typical commercial or 
hospital environment.