9-35 
those clotting curves with high noise level in the absorbance along the curve. If this is the case, CND 
will be reported as a result (see Section 9.2.2.5). 
 
 
Figure 9.21. Graphic Description of the “Maximum Acceleration for Extrinsic Pathway” Algorithm 
Derived Fibrinogen 
This  algorithm  estimates  the  fibrinogen  concentration  in  the  sample  from  the  PT  primary  curve  by 
calculating the absorbance increase (mE) between the beginning and the end of clot formation. 
Point 1 is the point, on the primary curve, where the first derivative is maximum. 
Point 2 and Point 3 are the closest points under and over Point 1 where the first derivative has a 
value that is the 20% and 10% of its maximum, respectively. 
The  absorbance  increase  between  Point  2  and  Point  3  gives  the  estimation  of  fibrinogen 
concentration in mE
 
 