QManager: Other Options 
 
9-36 
Fibrinogen Clauss 
This algorithm estimates the clotting time (in seconds) of the sample by calculating the 50% of the 
absorbance increase (mE) between the beginning and the end of clot formation.  
Point 1 is the point, on the primary curve, where the first derivative is maximum. 
Point 2 and Point 3 are the closest points under and over Point 1 where the first derivative has a 
value that is the 20% of its maximum. 
The  half  of  the  difference  between  the  Point  2  and  Point  3  returns  the  clotting  time  result  (in 
seconds). 
 
Figure 9.23. Graphic Description of the “Fibrinogen Clauss” Algorithm 
Time to Absorbance Increment 
The clotting time (in seconds) corresponds to the time in which the absorbance has increased 40 mE 
from the initial absorbance (A
0