Woodsonix Limited WSSB800TX User Manual

Page of 7
 
 
1. Overview 
The SB800 is an ultrasonic test unit for use in making single pulse transmission 
time measurements in solid samples.  The SB800 consists of three units: 
 
 
A transmitter probe (
Red
 
A receiver probe (
Blue
 
A display module (
Silver
) where readings are given in microseconds (μS) 
 
The unit operates by the transmitter probe producing an acoustic pulse at the tip 
of the probe and timing the subsequent interval until the pulse is successfully 
detected at the receiver probe.  The transmission time will depend on the 
propagation within the sample and the thickness of the sample.   
 
The three units are switched on automatically when they are removed from the 
carry case.  Note that the carry case contains magnetic material that may damage 
sensitive items, such as credit card strips, that are held in close proximity. 
 
2. High-Voltage Warning! 
The transmitter probe has a high voltage present internally.  This can cause a 
painful electric shock, and for this reason we recommend that you return faulty 
units to Woodsonix or its authorized dealers for servicing.  
 
The unit should not be operated with the probes immersed in any liquid. 
 
 
3. Intellectual Property 
All intellectual property relating to the concept, design and implementation of the 
SB800 hardware and software belongs to Woodsonix Limited.  Copyright, patent, 
and other infringements will be dealt with aggressively. 
 
 
4. Operating Instructions 
The units are switched on automatically by removing them from their carry case.  
In order to prolong battery life, the units should be turned off (by replacing the 
base unit and probes in the case) whenever they are not in use.  
 
4.1 Unit Identification 
The probes are colour coded. The 
transmitter probe has a red coloured body
 and 
the receiver probe is blue
.  Readings are shown on the display unit's screen.  A 
sample screenshot is shown in figure 1. 
 
 
Figure 1. Display-unit screenshot 
 
4.2 Displayed Measurement 
With the probe tips firmly applied to a solid sample, a reading of the transmission 
time in microseconds should be displayed.  If no measurement is displayed this 
can indicate that the attenuation in the sample is too great, the transmission time is 
greater than 2000 microseconds, or the unit is faulty (check batteries).  Figure 2 
shows the display location of the reading. 
 
 
Figure 2. Reading portion of the display 
 
With  the  display  unit  turned  on  and  the  probes  held  apart,  the  screen  should 
show: ‘----’ instead of a reading, indicating that no measurement is being made.  
If the display shows a reading in this situation this may indicate a faulty/damaged 
probe or low battery voltage.