Woodsonix Limited WSSB800TX User Manual

Page of 7
 
 
4.3 Previous Reading 
The number in the top right of the display-module screen represents the most 
recent good reading.  Its display position is shown in figure 3. 
 
 
Figure 3. Previous reading portion of the display 
 
 
4.4 Battery-Level Indicators 
There are three battery-level indicators on the right of the display-module screen 
as shown in figure 4.  See section 6 for more section for more information on 
battery replacement. 
 
 
Figure 4. Battery life indicator portion of the display 
 
The unit is carefully calibrated after manufacture, and is rechecked during the 
quality assurance stage.  In order to test this calibration, and the correct operation 
of the unit, press the two probe tips together; a reading of ‘0’ should be displayed. 
 
 
5. Serial Number Identification 
Each SB800 component is uniquely identified with a serial number. In addition to 
a physical number, an electronic serial number is stored in each component and 
these are shown on the screen when the display module is turned on.  Note that 
the displayed probe serial numbers are those in use when the display unit was 
turned off.  Therefore, the first time replacement probes are used, the displayed 
serial numbers will not match the actual probes. 
 
 
6. Battery Replacement 
Each battery indicator is shown as a solid rectangle when the batteries are fully 
charged.  An empty, or near empty, rectangle indicates that the batteries in the 
associated unit should be changed.  The letter beside each battery tells which unit 
the battery reading is for: 
 
 
R (
Red
) for the transmitter probe 
 
B (
Blue
) for the receiver probe  
 
S (
Silver
) for the display module 
 
The unit will continue to give accurate timing measurements when the battery 
voltage is low, as long as the transmit pulses continue, but the sensitivity will be 
reduced.  Ensure that the batteries are inserted correctly and never mix old 
batteries with new ones.  For maximum battery life, alkaline batteries should be 
used. 
 
6.1 Replacement of Probe Batteries 
Unscrew the cap at the rear of the probe by turning it anticlockwise.  The battery 
holder can be slid out and the batteries changed.  Ensure that the terminals on the 
end of the battery holder point towards the probe tip when it is re-inserted, and 
that they line up the metal contacts in the probe.  Figure 5 shows the probes with 
unscrewed caps. 
 
6.2 Replacement of Display-Unit Batteries 
Remove the four screws on the rear of the display unit to open up the battery 
cavity.  Replace the batteries in the battery holder.  Do not remove the plate that 
covers the front of the display unit, as it seals the electronic and microprocessor 
components from environmental damage.  Figure 5 shows the display unit with 
the battery cavity visible.