Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
46
System settings
5.5.7 Safety Test
This function is required for the annual safety check. Refer to the chapter “Maintenance, 
repair and shipment” on page 86. This function is used to close the output relays which 
disables the electrosurgical generator for normal operation. Closed output relays are needed 
to perform measurements.
 - The default setting is [Relays Off]:
The relays are open and the electrosurgical generator is enabled for normal operation.
 - The setting [Relays On]:
The relays are closed. The electrosurgical generator is disabled for normal operation.
As soon as the user exits this menu function, the relay setting is automatically set to default, 
i.e. normal operation of the electrosurgical generator.
To open or close the output relays proceed as follows:
1.  Tap [MENU] on the All Screen.
The screen [Select Menu] is displayed.
2.  Tap the arrow button [Next] to browse to the next page of the screen [Select Menu].
3.  Tap [Safety Test].
 - The screen [Safety Test] is displayed.
 - The current setting is highlighted.
4.  Tap [Relays On] or [Relays Off].
The new setting is highlighted.
5.  Tap [Cancel] to return to the screen [Select Menu].
5.5.8 Bipolar Cut Setup
The electrosurgical generator provides the options to switch the bipolar cutting modes on or 
off. If the bipolar cutting modes are switched on, the BIPOLAR pane is divided into a yellow 
button for cutting and a blue button for coagulation. The user can access and activate all 
bipolar cutting modes.
As a default setting the cutting modes for the BIPOLAR socket are switched off, i.e. only the 
blue coagulation button is displayed in the BIPOLAR pane. It is impossible to access and 
activate any bipolar cutting mode.
To switch the bipolar cutting modes on or off proceed as follows:
1.  Tap [MENU] on the All Screen.
The screen [Select Menu] is displayed.
2.  Tap the arrow button [Next] to browse to the next page of the screen [Select Menu].
3.  Tap [Bipolar Cut Setup].
 - The screen [Bipolar Cut Setup] is displayed.
 - The current setting is highlighted.
4.  Tap [On] or [Off].
The new setting is highlighted.
5.  Tap [OK] to return to the screen [Select Menu].
5.5.9 Service
This function is not intended for customer use. It is password protected and only accessible 
for authorized service centers.