Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
47
System settings
5.5.10 Volume control
WARNING
The output tones play an important role to notice the output power when the electrosurgical 
generator is activated. If the output tones are inaudible, then activation may not be 
recognized by the user. This could result in injury to the patient.
 
• Make sure that the output tones are at an audible volume level.
 
• Adjust the volume level of the output tones to the ambient noises.
The volume of the output tone can be adjusted either with the volume control on the rear 
panel of the electrosurgical generator or via touch screen. However, during activation, it is 
impossible to enter the screen [Select Menu]. Therefore, the volume needs to be adjusted 
with the volume control at the rear panel during activation.
The volume settings range from 1 to 10. The default setting is 7.
To set the volume for the output tone via the touch screen proceed as follows:
1.  Tap [MENU] on the All Screen.
The screen [Select Menu] is displayed.
Figure 5.7  Volume control
2.  In the volume pane tap [+] to increase the volume. Tap [−] to decrease the volume of the 
output tone.
Volume of error tones
The volume of the error tones (low, medium and high priority errors) are not adjustable. 
Refer to the chapter “Troubleshooting” on page 78 and to the sections “Alarm system” on 
page 97 and “Output tone information” on page 98in the Appendix A.
5.5.11 Brightness control
Use this function to control the brightness of the touch screen. The brightness settings range 
from 1 to 10. The default setting is 5.
To set the brightness proceed as follows:
1.  Tap [MENU] on the All Screen.
The screen [Select Menu] is displayed.