Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
50
Connecting HF instruments
CAUTION
No connection of a 2-pin plug with 4 mm pin diameters
Figure 6.2  CAUTION - Do not connect a 2-pin plug with 4 mm pin diameters to the 
MONOPOLAR socket
Connecting a 2-pin plug with pin diameters of 4 mm, e.g. hand switch activated monopolar 
forceps, to the MONOPOLAR output socket can damage the electrosurgical generator.
 
• Do not connect monopolar HF instruments with  2-pin plugs.
 
• Only connect plugs to the MONOPOLAR output socket as specifi ed in the section 
"Description of the output sockets" on page 50.
CAUTION
No connection of plugs with single connecting pins
If bipolar HF instruments are used, that have single plugs for each connecting pin, then 
there is the risk that the single plugs are connected accidentally to the wrong receptacles. 
This could result in damage of the electrosurgical generator.
 
• Only use bipolar HF instruments that have plugs as described in the section “BIPOLAR 
output socket specifi cations (applied part)” on page 51.
6.2  Description of the output sockets
Monopolar and bipolar instruments
The electrosurgical generator can be used with monopolar cutting and coagulation 
instruments as well as with bipolar cutting and coagulation instruments. The HF instruments 
must have an appropriate plug to connect either into the MONOPOLAR socket or the 
BIPOLAR socket, see socket descriptions below.
Maximum output voltage
When connecting HF instruments avoid generator settings, e.g. mode, effect, where the 
maximum output voltage exceeds the allowed peak voltage of the HF instrument. Refer to 
the sections about mode and output characteristics in the “Appendix A” on page 97. Also 
refer to the instructions for use of the HF instrument.