Connecting HF instruments
6 Connecting HF instruments
6.1 Safety information for connecting HF instruments
WARNING
Proper connection of the plug
If the HF instrument's plug is not connected properly, then the energy output may not be 
suffi cient. There is the risk that the HF instrument cuts tissue mechanically. This could result 
in bleeding and perforation of the tissue.
 
• Properly connect the plug of the HF instrument to the corresponding output socket of the 
electrosurgical generator as described in the following sections.
CAUTION
Correct connection of a monopolar 1-pin plug with 4 mm pin diameter
Figure 6.1  CAUTION - Correct connection of a monopolar 1-pin plug with a 4 mm 
pin diameter
If a 1-pin plug with a pin diameter of 4 mm is connected to any other receptacle than the 
right-hand receptacle of the MONOPOLAR output socket, then the output socket may be 
destroyed during activation.
 
• Connect a 1-pin plug with a 4 mm pin diameter only the right-hand receptacle as shown in 
the fi gure above.