Olympus Winter & Ibe GmbH ESG200 User Manual

Page of 110
80
Troubleshooting
Problem
Cause
Remedy
Foot switch pedal or 
hand switch of the HF 
instrument is pressed 
and the output tone 
sounds but no output 
power is delivered.
The foot switch is assigned to another 
output socket.
Check the correct assignment of the 
foot switch. Refer to the section “Foot 
switch assignment and Autostart” on 
page 41.
Improper connection of the HF 
instrument plug with the output socket 
of the electrosurgical generator.
Check the plug of the HF instrument 
for correct connection.
Malfunction of the electrosurgical 
generator.
Do not use this electrosurgical 
generator.
Contact Olympus.
No output power 
is delivered when 
an RFITT mode is 
selected. The end-
of-procedure signal 
sounds immediately 
after activation.
The electrode has no contact with the 
tissue.
Check that the electrode has contact 
with the tissue.
Improper connection of the HF 
instrument's plug with the output 
socket of the electrosurgical 
generator.
Check the plug of the HF instrument 
for correct connection.
The HF instrument or its connection 
cable is damaged.
Replace the HF instrument or its 
connection cable.
Malfunction of the electrosurgical 
generator.
Do not use this electrosurgical 
generator.
Contact Olympus.
It is impossible to stop 
the output power of 
the electrosurgical 
generator.
Autostart is assigned to the currently 
used output socket and both 
electrodes touch the tissue.
Remove the electrodes of the HF 
instrument from the tissue.
Malfunction of the foot switch or hand 
switch.
Immediately switch off the 
electrosurgical generator by pressing 
the power switch. Replace the foot 
switch or the HF instrument with hand 
switch.
Malfunction of the electrosurgical 
generator.
Do not use this electrosurgical 
generator.
Contact Olympus.
Its impossible to switch 
off the electrosurgical 
generator.
Malfunction of the electrosurgical 
generator.
Disconnect the cord set plug from 
the socket on the rear panel of the 
electrosurgical generator and/or from 
the grounded mains electricity socket.
Contact Olympus.
Do not use this electrosurgical 
generator.
Table: Troubleshooting – General problems
12.2 Error messages
The electrosurgical generator is equipped with an intelligent alarm system. The alarm 
system determines alarm conditions on the base of multiple variables. 
DRAFT 
– 