Troubleshooting
3) Remedial actions
4) Confirmation button
If the cause that released the alarm is cleared, then the error message is closed
automatically after approximately 10 seconds. If the cause of the alarm is not cleared, then
the error message stays on the touch screen.
• To clear the alarm, follow the instructions that are given in the error message.
It is possible to tap [OK] in the error message to close the error message faster.
12.2.1 Overview of error messages
The following table gives an overview of all error messages that possibly can be displayed.
The error messages are sorted numerically according to the error code.
Error code Message
Cause
Remedial action
E0001
Open circuit
1. Check if the electrodes
of the instrument have
proper tissue contact.
2. If the problem persists,
replace the instrument.
3. If the problem persists,
contact Olympus.
The electrodes of the HF
instrument have no tissue
contact.
Ensure that the electrodes
of the HF instrument have
proper tissue contact.
Malfunction of the HF
instrument or its connection
cable.
Replace the HF instrument
or its connection cable.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0002
Short circuit
1. Ensure the instrument
electrodes do not come
into contact with each
other.
2. If the problem persists,
replace the instrument.
3. If the problem persists,
contact Olympus.
The electrodes of the HF
instrument touch each other.
Ensure that the electrodes
of the HF instrument do not
touch each other.
Malfunction of the HF
instrument or its connection
cable.
Replace the HF instrument
or its connection cable.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0012
Adjustment incomplete
If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0016
Burn-in incomplete
If the problem persists,
contact Olympus.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.
E0019
Connection of Footswitch
1. You can connect a double
pedal footswitch only.
2. If the problem persists,
contact Olympus.
A single pedal foot switch is
connected.
Disconnect the single pedal
foot switch.
Only connect the supplied
double pedal foot switch
(WB50402W).
Malfunction of the foot
switch.
Replace the foot switch.
Malfunction of the
electrosurgical generator.
Do not use this
electrosurgical generator.
Contact Olympus.