2/ DESCRIPTION
See descriptive diagram  at the end  of the User Manual.
Passage of the conductor
Jaws
Protective non-slip guard
Automatic zero DC button
Red fault light (over range / incorrect zero adjustment)
Green light indicating correct power supply
3-position sliding switch (off / selection of 1 mV/A or 10 mV/A ranges)
Hand-held parts
Fitted lead 1.5 m
BNC plug
3/ OPERATING PROCEDURE
3.1/ SWITCHING ON
Set the sliding switch 
 to the appropriate position, 40 A range  (sensitivity 10 mV/A) or 400 A
range (sensitivity 1 mV/A). Correct operation is indicated by a green light 
 indicating that
the battery is in good condition.
After approximately 10 minutes of operation of the clamp without manipulation of the
control buttons, the power supply cuts off automatically (see “ Auto off ” below).
If this green indicator does not come on when the clamp is switched on, or goes out
before it has operated for 10 minutes, it is necessary to replace the battery (see
MAINTENANCE chapter).
3.2/ DC ZERO ADJUSTMENT
Ensure that the jaws of the clamp are correctly closed and that they do not enclose any
conductor. Connect the clamp to your measurement instrument. Press the auto zero
button 
. The red light 
 comes on for approximately three seconds to indicate that the
instrument is on zero calibration. If zero can not be obtained, this indicator light remains
lit to indicate the fault.
3.3/ MEASUREMENT
After having switched on the clamp, connected it to the measurement instrument on the
appropriate range, and followed the auto zero procedure (see the two paragraphs above),
clamp the conductor to be measured 
 between the jaws of the clamp.
On DC current measurement, ensure that the arrow located on the external edge of
the jaws 
 corresponds to the direction of the current flowing in the conductor
(source 
 receiver).