Chauvin Arnoux P01120072Clip-on ammeter adapter 39 mm P01120072 User Manual
Product codes
P01120072
1 2
2/ DESCRIPTION
See descriptive diagram at the end of the User Manual.
➀
Passage of the conductor
➁
Jaws
➂
Protective non-slip guard
➃
Automatic zero DC button
➄
Red fault light (over range / incorrect zero adjustment)
➅
Green light indicating correct power supply
➆
3-position sliding switch (off / selection of 1 mV/A or 10 mV/A ranges)
➇
Hand-held parts
➈
Fitted lead 1.5 m
➉
BNC plug
3/ OPERATING PROCEDURE
3.1/ SWITCHING ON
Set the sliding switch
Set the sliding switch
➆
to the appropriate position, 40 A range (sensitivity 10 mV/A) or 400 A
range (sensitivity 1 mV/A). Correct operation is indicated by a green light
➅
indicating that
the battery is in good condition.
After approximately 10 minutes of operation of the clamp without manipulation of the
control buttons, the power supply cuts off automatically (see “ Auto off ” below).
If this green indicator does not come on when the clamp is switched on, or goes out
before it has operated for 10 minutes, it is necessary to replace the battery (see
MAINTENANCE chapter).
After approximately 10 minutes of operation of the clamp without manipulation of the
control buttons, the power supply cuts off automatically (see “ Auto off ” below).
If this green indicator does not come on when the clamp is switched on, or goes out
before it has operated for 10 minutes, it is necessary to replace the battery (see
MAINTENANCE chapter).
3.2/ DC ZERO ADJUSTMENT
Ensure that the jaws of the clamp are correctly closed and that they do not enclose any
conductor. Connect the clamp to your measurement instrument. Press the auto zero
button
Ensure that the jaws of the clamp are correctly closed and that they do not enclose any
conductor. Connect the clamp to your measurement instrument. Press the auto zero
button
➃
. The red light
➄
comes on for approximately three seconds to indicate that the
instrument is on zero calibration. If zero can not be obtained, this indicator light remains
lit to indicate the fault.
lit to indicate the fault.
3.3/ MEASUREMENT
After having switched on the clamp, connected it to the measurement instrument on the
appropriate range, and followed the auto zero procedure (see the two paragraphs above),
clamp the conductor to be measured
After having switched on the clamp, connected it to the measurement instrument on the
appropriate range, and followed the auto zero procedure (see the two paragraphs above),
clamp the conductor to be measured
➀
between the jaws of the clamp.
On DC current measurement, ensure that the arrow located on the external edge of
the jaws
the jaws
➁
corresponds to the direction of the current flowing in the conductor
(source
⇒
receiver).