Chauvin Arnoux P01120072Clip-on ammeter adapter 39 mm P01120072 User Manual

Product codes
P01120072
Page of 52
1 3
3.4/ OVERLOAD INDICATION
Detection of overload of the range of the clamp is indicated by the red light 
. This
indicator flashes for a peak current greater than 60 A on the 40 A range (10 mV/A)  or
600 A on the 400 A range (1 mV/A).
3.5/ AUTO OFF
The clamp has an Auto Off feature which switches off 10 minutes after the clamp has been
switched on, without the controls being used.
When the clamp is switched off by this automatic function, the switch 
 must first be set to the
OFF position before being switched on again.
This function can be overridden by the user when switching on. Simply press the auto zero
button 
 at the same time as setting the switch 
 from the OFF position to the 1 mV/A or
10 mV/A position. If the green indicator 
 flashes whilst the zero reset button is being pressed,
this indicates that the auto off function has been inhibited.
4/ SPECIFICATIONS
Ranges
input/output ratio
measurement extent
AC rms
peak max
DC
40 A
10 mV/A
0.2...40 A
0.2...60 A
0.4...60 A
400 A
1 mV/A
0.5...400 A
0.5...600 A
0.5...600 A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: Earth’s DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current 
 65 Hz
- Impedance of the measurement instrument : 
 1 M
Ω
 and 
 100 pF
4.2/ OPERATING CONDITIONS
The instrument must be used in the following conditions to satisfy the safety of the user and
the metrological performance:
- Use indoors
- Working altitude: 
 2000 m
- Transportation altitude: 
12 000 m
- Environmental conditions: see graph below