Basic Operation 
DSO-1062D Digital Oscilloscope User Manual                                             34 
to a known setup. 
See below for waveform menus.   
 
Note: The oscilloscope will save the current settings if you wait 5  seconds after the last 
modification, and it will recall these settings the next time you power on the oscilloscope. 
5.5.2 MEASURE 
Push the MEASURE button to perform auto measurements. There are 11 types of measurements 
and up to 8 can be displayed at a time.   
Turn the V0 knob to select an unspecified option. Press V0 or F6 when the red arrow icon stops on 
it. Then the following menu appears.   
Options 
Settings 
Comments 
Source 
CH1   
CH2 
Select the measure source. 
Measurement 
Type 
Frequency 
Calculate the waveform frequency by measuring the first cycle. 
 
Period 
Calculate the time of the first cycle. 
 
Mean 
Calculate the arithmetic mean voltage over the entire record. 
 
Pk-Pk 
Calculate the absolute difference between the greatest and the 
smallest peaks of the entire waveform. 
 
Cyc RMS 
Calculate the actual RMS measurement of the first complete 
cycle of the waveform. 
 
Min 
Examine the waveform record  of all points in the current 
window and display the minimum value.   
 
Max 
Examine  the waveform record of all points in the current 
window and display the maximum value. 
 
Rise Time 
Measure the time between 10% and 90% of the first rising 
edge of the waveform. 
 
Fall Time 
Measure the time between 90% and 10% of the first falling 
edge of the waveform. 
 
Positive Width 
Measure the time between the first rising edge and the next 
falling edge at the waveform 50% level. 
 
Negative Width 
Measure the time between the first falling edge and the next 
rising edge at the waveform 50% level. 
 
Off 
Do not take any measurement. 
Use the knob V0 or the functional keys F3, F4 to select the type of measurement. 
The white waveforms on the menu is 
the recalled RefA waveform 
At most 9 groups of 
setups can be stored