Basic Operation 
DSO-1062D Digital Oscilloscope User Manual                                             36 
default). 
 
Peak Detect: Use this mode to detect glitches within 10ns and to limit the possibility of aliasing. 
This mode is valid at the SEC/DIV setting of 4µs/div or slower. Once you set the SEC/DIV setting 
to 4µs/div or faster, the acquisition mode will change to Normal because the sample rate is fast 
enough that Peak Detect is unnecessary. The oscilloscope does not display a message to tell you 
that the mode has been changed to Normal. 
Average:  Use this mode to reduce random or uncorrelated noise in the signal to be displayed. 
Acquire data in Normal  mode and then average a great number of waveforms. Choose the 
number of acquisitions (4, 16, 64 or 128) to average for the waveform. 
Stopping the Acquisition: When you are running the acquisition, the waveform display is live. 
Stop the acquisition (press the RUN/STOP button) to freeze the display. In either mode, the 
waveform display can be scaled or positioned by vertical and horizontal controls. 
Equivalent Acquisition:  Just  repeat the Normal  acquisition. Use this mode to take a specific 
observation  on repeatedly displayed periodic signals. You can get a resolution of 40ps, i.e. 
25GSa/s sample rate, which is much higher than that obtained in real-time acquisition. 
The acquisition principle is as follows. 
 
 
As shown above, acquire input signals (cycle repeatable) for more than once at a slow sample rate, 
arrange the sample points by the time they appear, then recover waveforms.   
 
Sample Points 
 
Normal Acquisition Intervals
 
4  5 
9  10 
Normal Mode Acquires a Single Sample Point in Each Interval
 
Input repeated signals 
First Acquisition 
Second Acquisition 
Third Acquisition 
Fourth Acquisition