Voltcraft DSO-1062D 2-Channel Oscilloscope, Digital Storage Oscilloscope DSO-1062D User Manual

Product codes
DSO-1062D
Page of 84
Basic Operation 
DSO-1062D Digital Oscilloscope User Manual                                             40 
Trigger Video Standard 
Adjusted 
Vertical Bandwidth 
Full 
Vertical Coupling 
DC (if GND was chosen before); AC for the video signal; otherwise, 
unchanged 
VOLTS/DIV 
Adjusted 
The Autoset function examines all channels for signals and displays corresponding waveforms. 
Autoset determines the trigger source according to the following conditions. 
  If multiply channels get signals, the oscilloscope will use the channel with the lowest 
frequency signal as the trigger source. 
  If no signals are found, the oscilloscope will use the lowest-numbered channel displayed 
in Autoset as the trigger source.   
  If no signals are found and no channels are displayed, the oscilloscope will display and 
use Channel 1 as the trigger source. 
Sine Wave: 
When you use the Autoset function and the oscilloscope determines that the signal is similar to a 
sine wave, the oscilloscope displays the following options.   
Sine Wave Options 
Details 
Multi-cycle Sine   
Display multiple cycles that have appropriate vertical and 
horizontal scales. 
Single-cycle Sine 
Set the horizontal scale to display about one cycle of the waveform. 
FFT 
Convert the input time-domain signal to its frequency components 
and display the result as a graph of frequency versus amplitude 
(spectrum). Since it is a math calculation, see 
 for more information. 
Undo Setup 
Let the oscilloscope recall the previous setup. 
Square Wave or Pulse: 
When you use the Autoset function and the oscilloscope determines that the signal is similar to a 
square wave or pulse, the oscilloscope displays the following options.   
Square Wave Options 
Details 
Multi-cycle Square 
Display multiple cycles that have appropriate vertical and 
horizontal scales. 
Single-cycle Square 
Set the horizontal scale to display about one cycle of the waveform. 
The oscilloscope displays Min, Mean and Positive Width automatic 
measurements.   
Rising Edge 
Display the rising edge. 
Falling Edge 
Display the falling edge. 
Undo Setup 
Let the oscilloscope recall the previous setup. 
5.6.2 Help 
Push the HELP button to display the Help menu which provides  topics covering  all  the menu 
options and controls of the oscilloscope. For detailed information about the Help system, refer to 
.