DSO-1062D Digital Oscilloscope User Manual                                             41 
5.6.3 Default Setup 
When you push the DEFAULT SETUP button, the oscilloscope will display the CH1 waveform and 
remove all the others. When you are at the default setup, press F1 to Undo  Preset.  Then the 
oscilloscope returns to the status before default setup. The table below gives the options, buttons 
and controls that change settings at default setup. 
Menu or System 
Option, Button or Knob 
Default Setting 
Acquire 
(Three mode options) 
Normal 
Averages 
16 
Run/Stop 
Run 
Cursor 
Type 
Off 
Source 
CH1 
Horizontal (amplitude) 
±3.2div 
Vertical (time) 
±4div 
Display 
Type 
Vectors 
Persist 
Off 
Format 
YT 
Horizontal 
Window Mode 
Single-window 
Trigger Knob 
Level 
Position 
0.00s 
SEC/DIV 
2
00μs 
Math 
Operation 
— 
Source 
CH1-CH2 
Position 
0div   
FFT 
Vertical Scale 
20dB 
FFT Operation   
 
Source 
CH1 
Window 
Hanning 
FFT Zoom 
X1 
Measure 
Source 
CH1 
Type 
None 
Trigger (Edge)
 
Type 
Edge 
Source 
CH1 
Slope 
Rising 
Mode 
Auto 
Coupling 
DC 
Level 
0.00v 
Trigger (Video) 
Polarity 
Normal 
Sync 
All lines 
Standard 
NTSC 
Trigger (Pulse) 
When 
Set Pulse Width 
1.00ms 
Polarity 
Positive 
Mode 
Auto