  User‟s Guide for DS1000B Series 
 
7.  The signal is displayed as ladder like waveform. 
(1)  The time base setting maybe is too slow. Turn the horizontal 
  knob 
to increase horizontal resolution to improve the display.   
(2)  Maybe the display Type is set to “Vectors”, and set it to “Dots” mode to improve 
the display. 
 
8.  Unable to connect to PC or the PictBridge printer by USB
 
(1)  When connected to PC, if the message “Unknown device” appears or prompting 
scanner or camera was found. Press Utility to choose the I/O Setting menu, you 
may  see  the  USB  Device  menu  has  been  set  to  be  “PictBridge”,  you  should 
switch it to “Auto Detect” or “Computer”. If necessary, restart the oscilloscope. 
(2)  If connecting to a PictBridge printer or the printing is unsuccessful, maybe the 
USB Device menu has been set to “Computer”. You should switch it to be “Auto 
Detect” or “PictBridge”. If necessary, restart the oscilloscope.