Greisinger EASYLOG 80 CL Humidity: 0 - 100 % r.H · Temperature: -25 to +60 °C · Pressure: 300 - 1100 hPa Temperature Dat 603414 Data Sheet

Product codes
603414
Page of 12
E40.0.01.6C-02 Operating 
Manual 
EASY
LOG
 80 CL 
Page 9 of 12 
9.3 System messages 
... 
°C
%
 
 
BATTERY
• display of the  BAT-arrow left hand down: 
The 
EASY
LOG
 battery is almost spent and needs to be replaced.
The data recording is still active. 
• permanent display 
BAT
  The battery is spent and the recording has been automatically 
stopped (even if at EASYBus-operation).  
It is still partly possible to read the measured values and to  
read-out the saved logger data. 
  => Please return the logger to the manufacturer. 
 
ERROR 
7
The 
EASY
LOG
 has detected a system error. 
• remedy:   Reset the error via GSOFT 40K. 
  If the error message is displayed furthermore please  
  send the logger to the manufacturer for repair. 
 
 
9.4  Alarm and error messages 
The following messages correspond to a particular measuring channel and are dis-
played accordingly. 
 
ALARM LOW
The measured value is below the selected min-alarm limit. 
This and the current value is displayed alternately.  
 
ALARM HIGH
The measured value is above the selected max-alarm limit. 
This and the current value is displayed alternately.
 
 
ERROR 
1
The measuring range of this channel is exceeded. 
 
ERROR 
2
The measuring range of this channel is under-run. 
 
ERROR 
9
The sensor for this channel provides invalid values. 
 
ERROR 
11
The variable could not be calculated, because the measuring  
range has been exceeded or a measured value, that is needed for
the calculation, is not available or faulty. 
 
Note: 
A description of the possible reasons for that error messages is in 
chapter 12 (FAQ).