How to Operate 
15 
External Sensor 
In addition to the integrated vibration sensor, an optional 
external sensor connects to the Meter. The connector type 
for the external sensor is a Subminiature version B (SMB). 
Figure 3 shows how to connect an external sensor to the 
Meter. 
Note 
Fluke supports, but does not provide, external 
sensors. 
 
gqi05.eps 
Figure 3. External Sensor Connection 
Note 
High frequency measurement (Crest Factor+) and 
Temperature measurement automatically turn off 
when an external sensor is connected to the 
Meter. 
To connect: 
1.  Open the connector cover and push the external 
sensor into position. 
After you connect the external sensor, a pop-up 
message opens on the Meter display and then 
opens the Enter Sensitivity menu. 
 
gqi14.eps 
Note 
The sensitivity must be set in mV/g units. 
2. Push 