Line-Resistance at version for Pt 100-sensors (RTD):
Select menu item with Up/Down until display L-A and programmed parameter alternate.
Change into parametrizing mode with Set and do the line-compensation with Up/Down (enter
value of line-resistance) or 3-wire (3-L).
Change into parametrizing mode with Set and do the line-compensation with Up/Down (enter
value of line-resistance) or 3-wire (3-L).
2-wire-configuration, compensation of cable-resistance:
Short-circuit the wires nearby the sensor and measure the resistance of the cable. Set
„LA“ to this value.
„LA“ to this value.
Measuring Unit (°C or °F) at version for Pt 100-sensors (RTD):
Select menu item with Up/Down until display Unit and programmed unit alternate.
Change into parametrizing mode with Set and select unit (°C or °F).
Change into parametrizing mode with Set and select unit (°C or °F).
Tips:
- After finishing one menu item it is switched automatically to the next one.
- When the right decimal point in the 7 segment display is on, the display mode has been left,
and the menu items can be chosen with up/down (menu mode).
- When the right decimal point blinks, you are in the parameter setting mode and can change
the setting with up/down.
- Long pushing on up/down speeds up the changes in the display.
- Pushing button up and down at the same time sets values to zero.
- With reset (press set/reset for 2s) the display mode can be reached from every position of the
- Pushing button up and down at the same time sets values to zero.
- With reset (press set/reset for 2s) the display mode can be reached from every position of the
parameter setting mode (the last selected value in is being stored).
- With InLo and InHi scaling is simplified when measuring-range and range of signal are
different.
Example: Display 0-500.0 at input-signal 10-90 V:
Example: Display 0-500.0 at input-signal 10-90 V:
• Connect to measuring input 100 V and select InPU = 100
• Set InLo to 10.0
• Set InHi to 100.0
• Set diLo to 0
• Set
• Set InLo to 10.0
• Set InHi to 100.0
• Set diLo to 0
• Set
diHi to 5000
• Set dP to 000.0