Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 38 
 
www.PressureSystems.com 
Command ‘C’— Sub-command Index 03: Abort Multi-Point Calibration 
This sub-command has no additional parameters.  
 
Command 
C 03” 
 
‘C’  is the command letter. 
‘ 03’ is the sub-command index (ii) for Abort
 
NOTE: all parameters are separated by a space. 
 
Response 
A” 
 
‘A’ is the acknowledge letter  
 
Description: Aborts the Multi-Point Calibration function, if  it is currently in progress. This  
 
 
  sub-command also restores the original “A/D samples for averaging” count to the  
 
 
  module that was in use before the calibration function was started.  
 
  It should be noted that executing the Configure & Start (‘00’) sub-command  
  again, after the calibration function has started collecting data (per Collect Data  
  
(‘01’) sub-commands), but before the final data are calculated (per Calculate &  
 Apply
 (‘02’) sub-command), will have the same affect as this Abort function. 
 
Example: 
 
● 
Abort the Multi-Point Calibration function previously started 
 
 
 
C 03” 
 
  Read 
response: 
 
 “A”