Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 55 
 
www.PressureSystems.com 
Command ‘c’— Sub-command Index 03: Clear Stream(s)
 
This sub-command is used to “undefine” any previously configured host stream in a module, 
one at a time, or all together. The sub-command’s format is:  
 
 
Command 
c 03 st” 
 
‘c’  is the command letter 
‘03’ is the sub-command index (‘ii’) for configuration 
st’ is the stream identifier character (1, 2, or 3 or 0=all streams)  
 
NOTE: all parameters are separated by a space
 character
 
Response 
A” 
‘A’ is the acknowledge letter 
 
Autonomous 
Packet 
none generated 
 
Description:  This sub-command clears (un-defines) the particular specified host stream (st=1-
3), or un-defines “all configured” host streams with a single command (st=0). 
Once cleared, a stream must be reconfigured before it can be started.  
Example: 
● 
 
Stop all the streams configured previously.  Then clear (un-define) only stream 3.   
 
 
Finally, resume the remaining defined streams 1 and 2: 
 
 
 
 “c 02 0”    
 
 
 “c 03 3”    
 
 
 “c 01 0”    
 
 
 
  Read 
response: 
 
 
 
“A” 
 
 
“A” 
 
“A”