3.2.3    Mounting a Specimen on the Type I Stage (Manual Stage) 
 
IMPORTANT:  The specimen base that comes with the system is round.    When loading an 
angular specimen, select an appropriate specimen base so that the entire 
specimen will fit on the base.    Movable range limits are established by restricting 
the operating range for the X and Y-axes to within the diameter of the specimen 
stand and by calculating the safe operating range for the Z and Tilt axes.   
For example, if a square specimen measuring 50 mm on the edge is affixed onto 
a 50mm-diameter specimen base and the specimen is rotated 45°, the size of 
the specimen in X and Y directions will be 70 mm, which means that the 
specimen cannot be protected safely if its size is set to 50 mm.    In such a 
situation, either use a 77-mm (3-inch) specimen base or specify a specimen 
setting (see 3.5.6.6) of 77 mm (3 inches) rather than 50 mm (2 inches). 
 
 
3.2.3.1    Setting the Specimen Stand 
 
Combine the adjustable screw, locking ring, and holder base, and screw the stub to the 
adjustable screw.    Tighten the locking ring so that the adjustable screw will not become loose. 
 
 
 
 
Fig. 3.2-1    Fixing the Specimen Stub 
 
 
3.2.3.2    Measuring and Setting the Specimen Height 
 
Loosen the locking ring for the specimen stub.    Use an adjustment screw to align the highest 
position of the specimen to the Standard (45 mm) mark on the height gauge, and fix the 
specimen using the locking ring so that the adjustment screw will not become loose.    The height 
of the specimen base from the bottom to the specimen surface is 45 mm. See Section 3.5.7.3 for 
further details. 
 