3.2  Mounting a Specimen 
3 - 6 
 
 
 
 
 
 
 
 
 
 
 
Fig. 3.2-2    Measuring the Specimen Height 
 
IMPORTANT:  If the specimen is tilted or the Stage Z is short, the specimen can touch the 
objective lens or the backscattered electron detector that is provided on the 
underside of the objective lens, which can result in damage to the specimen, 
objective lens or the backscattered electron detector. 
 
3.2.3.3    Moving the Stage to the Specimen Exchange Position 
 
Align the X, Y, R, T, and Z knobs of the specimen stage to the specimen exchange position (the 
specimen exchange position is indicated on the upper face of the specimen stage). 
 
Specimen replacement position 
X: 30mm 
Y: 20mm 
Z: 35mm 
 
R: 0° 
T: 0° 
 
 
 
 
Fig. 3.2-3    Type I Stage 
Specimen stub   
Specimen height (Standard) 
Level gauge 
Specimen holder 
Specimen 
Locking ring 
Adjusting screw 
R
T