Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
1.1  Specifications 
1 - 2 
1.1.4  Specimen Goniometer Stage 
 
(a) 
Type I (80 x 40 stage) 
Range of motion: 
X-axis: 0 to 80 mm 
 
Y-axis: 0 to 40 mm 
 
Z-axis (WD): 5 to 35 mm 
 
R-axis (rotation): 360°, continuous 
 
T-axis (tilt): -20° to 90° 
Drive method: 
Manual drive 
Specimen size observable over entire region: 106 mm diameter (also R) 
Maximum specimen size: 
200 mm diameter   
 
(b) 
Type II (100 x 50 stage)   
Range of motion: 
X-axis: 0 to 100 mm 
 
Y-axis: 0 to 50 mm 
 
Z-axis (WD): 5 to 65 mm 
 
R-axis (rotation): 360°, continuous   
 
T-axis (tilt): -20°to 90° 
Drive method: 
5-axis motor drive 
 
With eucentric rotation feature   
 
With image navigation feature   
Specimen size observable over entire region: 130 mm diameter (also R) 
Maximum specimen size: 
200 mm diameter   
 
 
1.1.5  Evacuation System 
 
Evacuation control: 
Fully automatic pneumatic valve method 
Vacuum gauge: 
Pirani gauge 
Attainable vacuum level: 
1.5 x 10
-3
 Pa 
Range of vacuum level settings:  6 to 270 Pa (22 stops) 
Vacuum pump: 
Turbo molecular pump (TMP) 210 L/s for high vacuum 1 ea. 
 
Rotary pump 135 L/min (50Hz), 162 L/min (60Hz) 
 
TMP back evacuation/low-vacuum settings 1 ea. 
Compressor: 
0.2 kW output, 1 ea. 
 
 
1.1.6  Imaging Function 
Display monitor:   
Still image display on the PC monitor (at full scanning speed) 
Observation monitor: 
18.1- or 19-inch LCD (number of desktop pixels: 1280 x 1024) 
Image display mode: 
Full-screen display (1280 x 960 pixels) 
 
Small-screen display (640 x 480 pixels) 
 
Dual-screen display (640 x 480 pixels, two panes), with 
signal-mixing feature