Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
2.3    Graphical User Interface (GUI) 
2 - 40 
2.3.7.2    Focusing Position (WD) WORKING DISTANCE Window 
 
This block allows you to set a focusing position. 
 
 
 
 
Fig. 2.3-26    Focusing Position (WD) Window 
 
(1)  WD setup/display window 
This window displays the current working distance (the focusing position calculated from the 
lens current). 
When a working distance value is selected from the pull-down list, focus point will be set to 
the specified working distance.    The box also indicates present focal length if varied with 
manual focusing.    This window can be used for focusing purposes when the Z-axis of the 
stage is moved or when it is necessary to set the WD accurately for EDX analysis purposes. 
 
(2)  WD micro-adjustment button 
By clicking the Up and Down arrows, you can adjust the WD in 0.1mm increments. 
 
(3)  WD setting slider 
The slider indicates the current WD.    Also, by moving the slider with the mouse to the right 
and left, you can change the WD.    By clicking the side of the slider, you can adjust the WD 
in 1 mm increments. 
 
(4)  Analysis position setting button 
This button can be used to set the WD at 10 mm, such as during an EDX analysis. 
This button sets a WD without the moving the stage itself. 
 
WD setup/display box 
 
WD micro adjustment button 
WD setting slider 
Analysis position setting button