Emerson HASXEE-IM-HS User Manual

Page of 502
Emerson Process Management GmbH & Co. OHG
5-22
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
Signal:
Comp1
OutRange:
0-20 mA
Low Scale:
0.00
Max Scale:
100.00
AutoScale:
Yes
FailMode:
Live
0/4 mA:
0.00
20 mA:
100.00
Hold:
No
5.7   Checking the Settings
"0/4  mA"  enables  to  finetune  the  analog 
output:  Set  "Signal"  to 
0  mA  and,  while 
measuring the output current, adjust it to the 
expected value.
Accepted range: 
-10,000 ... +10,000
"20 mA" enables to finetune the analog output: 
Set "Signal" to 
20 mA and while measuring 
the output current, adjust it to the expected 
value.
Accepted range: 
-10,000 ... +10,000
"Hold"  selects  the  output´s  behaviour  during 
calibrations. 
If set to 
Yes,
•  the analog output is fixed to the last measu-
red value;
•  concentration alarms, which may otherwise 
be triggered by the concentrations of the 
calibration gases, are supressed.
If set to 
No
•  the  analog  output  signal  always  corre-
sponds to the actual measured value during 
calibration; this may trigger alarms if limits 
are exceeded. 
Note!
This behaviour may be undesireable if e.g. 
the unit is connected to a data acquisition 
system.