Emerson HASXEE-IM-HS User Manual

Page of 502
Emerson Process Management GmbH & Co. OHG
5-24
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
04/2010
5.7   Checking the Settings
You may turn the alarm function 
On or Off  
separately for each channel ("Alarm Monitor").
It´s  also  possible  to  use  only  some  of  the 
thresholds. In this case, set the not used to a 
level outside the range limits (for this, "Lo" and 
"LoLo" support entering negative values).
In  case  of  an  alarm,  you  may  like  to  enter 
the STATES submenu, to check which one 
is triggered.
The most important settings have now been 
checked and the unit is configured to suite 
your needs.
A  backup  copy  of  these  configuration  data 
can now be saved.
Press 
left
 until SETUP and then open SAVE-
LOAD. 
5.7.6  Backup the Settings
This menu gives you the choice, to either
•  make a local backup to a protected memory 
area
•  restore the factory default settings,   or
•  make  a  backup  to  an  external  USB  de-
vice.
Ch1
Concentration Alarms
LoLo Alarm:
Off
Lo Alarm
On
Hi Alarm
Off
HiHi Alarm
Off
Concentration
75.000 ppm
Save-Load
Local Backup..
Factory Defaults..
USB Backup..
USB Firmware Update..