7.4.1   Preparing a Calibration
Now open the next menu to assign up to 4 
span gas valves to the selected channel: one 
for each range.
Note!
Depending  on  the  gas  analyzer  and  SHS 
configuration, is may be possible to assign a 
specific valve to multiple ranges.
Again, don´t forget to specify the individual 
purge times.
On  the  3
rd
  menu  page,  assign  a  blowback 
valve  for  the  selected  channel,  if  such  is 
installed.
Ch1
  
Valve assignment 2of3
Span1 valve:
V1
Purge time:
1 s
Span2 valve:
None
Purge time:
0 s
Span3 valve:
None
Purge time:
0 s
Span4 valve:
None
Purge time:
0 s
Ch1
  
Valve assignment 3of3
Blowback valve:
None
Purge time:
0 s
Multi-channel unit:
On  menu  1,  press 
left
  to  open  SELECT 
COMPONENT to change the settings for a 
different channel.
Note!
To check, if entries, made on menu pages 2 
& 3 are correct, go back to menu 1 and check 
"Correct assign".