Maintenance & Procedures
7.4.2 Manual Calibration
7.4.2 Manual Calibration
7.4.2.1 Manual Zero Calibration
Starting from the MEASUREMENT SCREEN
press
down
to open the MAIN MENU and
enter CONTROL.
To start a zero calibration select the first
line
:
Multi-channel unit:
Select the channel to be calibrated in
SELECT COMPONENT.
Before selecting any further
line make sure the required
calibration gas is applied and
flowing!
Supply all calibration gases
Supply all calibration gases
with the same flow as the samp-
le gas (recommeded approx. 1 l/
min), pressureless and utilizing
the correct gas fitting (
sect.
3.4).
Ensure the warm-up time after
Ensure the warm-up time after
switching on has elapsed!
Warm-up time is 15 to 50 minutes
depending on installed measu-
ring system and configuration!
The first line gives you the choice to cancel
the procedure now.
Select the second line to
start the calibra-
tion.
Control..
Zero calibration..
Span calibration..
Advanced calibration..
Apply gas..
Lock menus!
Acknowledgements
Pump 1:
Off
Pump 2:
Off
Ranges..
Ch1
Zero calibration
Cancel!
Start!
Zero gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!