Maintenance & Procedures
7.4.2   Manual Calibration
When finished, press 
left
 several times to re-
turn to SELECT COMPONENT (multi channel 
analyzer only), to perform a span calibration 
for another channel,
or 
press 
home
 to return to the MEASUREMENT 
SCREEN,  to  finish  with  manual  calibration 
procedures.
7.4.2.2  Manual Span Calibration
Multi-channel unit: 
Select the channel to be calibrated in 
SELECT COMPONENT.
Before selecting any further line 
make sure the required calibrati-
on gas is applied and flowing!
Span calibration offers the same options as 
Zero calibration, so for a detailled description 
Select "Span calibration.."
Span calibration
Cancel!
Start!
Span gas
0.000 ppm
Concentration
0.000 ppm
Flow
0.00 l/min
Status..
Results..
Restore!
Zero calibration..
Span calibration..
Advanced calibration..
Apply gas..
Lock menus!
Acknowledgements
Pump 1:
Off
Pump 2:
Off
Ranges..