Barco DICOM THEATER LITE MGP D5 User Manual

Page of 116
Using TheaterWatch
MGP D5 Dicom Theater Lite
91
3. A wizard starts, guiding you through the calibration 
procedure. Follow the guidelines in the wizard.
The projector calibration procedure is a combination of 
actions in TheaterWatch and projector OSD (on-screen 
display) manipulations.
To calibrate a display using the Barco LCD sensor:
1. If you have multiple displays of the same type in the system, 
consider if you wish to equalize the display you calibrate to 
a reference display.
Equalization means the luminance of the display you are 
calibrating will be matched to the luminance of the reference 
display.
If you wish to equalize the display, you must close the 
Display Settings dialog and make sure the Equalization 
option in the Application Settings dialog is set. See 
“Calibration tab” in the description of the Application 
Settings dialog.
2. If you have multiple projectors or displays connected (multi-
head system), select the display you wish to calibrate by 
clicking on the corresponding numbered icon in the Display 
Settings dialog.
3. Click the Calibrate button.
4. A message appears, showing the different calibration 
options.
Figure 34: Calibration choices