Using TheaterWatch
92
MGP D5 Dicom Theater Lite
5. Select “Using the BARCO LCD sensor” and click OK.
This option is grayed in case TheaterWatch does not find a 
connected Barco LCD sensor.
6. Follow the guidelines on the screen to complete the 
calibration successfully.
Note:
If the Equalization option is set in the Application Settings 
and TheaterWatch finds at least one other display of the 
same type that is calibrated, the “Match with reference” 
page appears during the calibration process.
Figure 35: Match with reference 
To equalize the display, check the option “Match this display 
with display....” and click Next. The display number that 
appears in this option, refers to the reference display.