Barco DICOM THEATER LITE MGP D5 User Manual

Page of 116
Using TheaterWatch
MGP D5 Dicom Theater Lite
93
To calibrate a display manually:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor 
2. In the Calibration dialog, check the option Visually and click 
the OK button.
The DICOM Optimizer dialog appears.
Figure 36: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference 
between the background of the image above the slider and 
the bitmap inside this image.
4. When done, click button Set point 1 of 20.
5. Repeat this procedure until the button is grayed. This 
indicates the last point is set.
6. Click OK to finish.
To reset to the latest calibration:
1. Follow step 2 and 3 of the calibration with Barco LCD sensor 
2. In the Calibration dialog, check the option Use the latest 
calibration.