Normal Operation 
38 
Series N5700 User’s Guide  
Normal Operation 
The power supply has two basic operating modes: constant voltage 
and constant current mode. In constant voltage mode, the power 
supply regulates the output voltage at the selected value, while the 
load current varies as required by the load. In constant current 
mode, the power supply regulates the output current at the selected 
value, while the voltage varies as required by the load. The mode in 
which the power supply operates at any given time depends on the 
voltage setting, current limit setting, and the load resistance.  
Constant Voltage Mode 
When the power supply is operating in constant voltage mode, the CV 
indicator on the front panel illuminates. 
Adjustment of the output voltage can be made when the output is 
enabled (On) or disabled (Off). When the output is enabled, simply 
rotate the voltage knob to program the output voltage.  
When the output is disabled, press the LIMIT button and then rotate 
the voltage knob. The DC VOLTS display will show the programmed 
voltage for 5 seconds after the adjustment has been completed and 
then indicate OFF. 
The voltage knob can be set to coarse or fine resolution. Press the 
FINE button to select finer resolution. The FINE indicator turns on. 
NOTE 
If you cannot adjust the voltage to the value that you desire, the power supply 
may be operating at its current limit. Check the load condition and the current 
limit setting. Also, the voltage cannot be programmed lower than about 5% 
above the UVL setting, or higher than about 5% below the OVP setting. 
Constant Current Mode  
When the power supply is operating in constant current mode, the 
CC indicator on the front panel illuminates. 
Adjustment of the output current limit can be made when the output 
is enabled (On) or disabled (Off). When the output is enabled and in 
constant current mode, simply rotate the current knob to program 
the current limit. If the output is in constant voltage mode, press the 
LIMIT button and then rotate the current knob. The DC AMPS display 
will show the programmed current for 5 seconds after the adjustment 
has been completed and then indicate the actual output current. 
When the output is disabled, press the LIMIT button and then rotate 
the current knob. The DC AMPS display will show the programmed 
current for 5 seconds after the adjustment has been completed and 
then go blank because the output is off. 
The current knob can be set to coarse or fine resolution. Press the 
FINE button to select finer resolution. The FINE indicator turns on.