Protection Functions 
Series N5700 User’s Guide 
39 
CV/CC Mode Crossover  
If the power supply is in constant voltage mode and the load current 
increases above the current limit setting, the power supply switches 
to constant current mode. If the load decreases below the current 
limit setting, the power supply switches to constant voltage mode. 
CV/CC Signal  
CAUTION 
Do not connect the CV/CC signal to a voltage source higher than 30VDC. 
Always connect the CV/CC signal to the voltage source with a series resistor to 
limit the sink current to less than 10mA. 
The CV/CC signal available on the J1 connector indicates the 
operating mode of the power supply. The CV/CC signal is an open 
collector output with a 30V parallel zener at J1 pin 13, referenced to 
common at J1 pin 12. J1 pin 12 is connected internally to the –S 
terminal. When the power supply operates in constant voltage mode, 
CV/CC output is open. When the power supply operates in constant 
current mode, CV/CC signal output is low (0 - 0.6V), with maximum 
10mA sink current. 
Protection Functions 
Over-Voltage Protection 
The over-voltage protection protects against over-voltage conditions 
on the output. If the output voltage attempts to exceed the 
programmed limit in response to an analog programming signal or in 
the event of a power supply failure, the over-voltage protection 
circuit will protect the load by disabling the output. The voltage is 
monitored at the sense terminals, thus providing the protection level 
directly at the load. Upon detection of an over-voltage condition, the 
output is disabled, the display shows OVP, the PROT indicator blinks, 
and OV is set in the Questionable Condition status register. 
Adjustment of the over-voltage setting can be made when the output 
is enabled (On) or disabled (Off). To set the OVP level, press the 
OVP/UVL button so that the display indicates OUP. The display will 
show the OVP setting. Rotate the voltage knob to adjust the OVP level. 
The display will show OVP and the setting value for another five 
seconds and then return to its previous state. 
The OVP settings are limited at the minimum level to approximately 
5% above the output voltage setting. Attempting to adjust the OVP 
below this limit will result in no response to the adjustment attempt. 
Refer to Appendix A for the maximum OVP settings. 
Use one of the following methods to reset the OVP circuit after it 
activates. If the condition that caused the over-voltage shutdown is 
still present, the OVP circuit will turn the output off again.