Output On/Off Control 
Series N5700 User’s Guide 
43 
Enable/Disable Terminals 
CAUTION 
To prevent possible damage to the unit, do not connect the Enable + or Enable – 
terminals to the positive or negative output terminals.  
Enable/Disable terminals are available on the J1 connector to enable 
or disable the power supply output. This function is level-triggered. 
Simply connect a switch or relay between J1 pins 1 and 14. This 
function is activated by SW1 setup switch 9.  
These pins disable the output when they are opened. When the 
output is disabled, the PROT indicator on the front panel will blink. 
To re-enable the output after it has shut down, you must short the 
Enable + and Enable – terminals. In Auto-Restart mode, operation 
resumes automatically. In Safe-Start mode the Enable/Disable 
function is latched. You must also press the OUT ON button or send 
an OUTPut:PROTection:CLEar command to resume operation. 
 
SW1 switch 9 
ENA+/ENA– pins 
Output 
Display 
Prot Indicator 
Down (default) 
Not active 
On 
Voltage/Current 
Off 
Up Opened 
Off 
ENA 
Blinking 
 Shorted 
On 
Voltage/Current 
Off 
Power Supply OK Signal 
The Power Supply OK signal on the J1 connector indicates a fault 
condition in the power supply. J1 pin 16 is a TTL output signal. Pins 
2 and 3, which are connected internally, are the signal common. All 
pins are optically isolated from the power supply output. With no 
fault, Power Supply OK is high, with a maximum source current of 
2mA. When a fault occurs, Power Supply OK is low, with a maximum 
sink current of 1mA. The following faults set this signal low: 
Over-voltage protection 
Enable/Disable signal true 
Over-current protection 
Shut Off signal true 
Over-temperature protection 
GPIB failure 
AC line failure 
Output turned off 
Safe-Start and Auto-Restart 
The power supply can be programmed to have either the last 
operating settings (Auto-Restart) or the reset settings (Safe-Start) 
apply at turn-on. Press and hold the OUT ON button to select between 
Safe-Start and Auto-Restart modes. The display continuously cycles 
between SAF and AUT every three seconds. Releasing the OUT ON 
button while one of the modes is displayed, selects that mode.  
In Safe-Start mode, the power supply turns on with the reset 
settings (see chapter 5 under “*RST”). The output is disabled and the 
output voltage and current are zero. This is the factory default.