Agilent Technologies N5700 User Manual

Page of 128
System Commands 
Series N5700 User’s Guide 
77 
System Commands 
System commands control system functions that are not directly 
related to output control, measurement, or status functions. Common 
commands are also used to control system functions. 
SYSTem:COMMunicate:RLSTate LOCal|REMote|RWLock  
SYSTem:COMMunicate:RLSTate? 
This command configures the remote/local state of the instrument 
according to the following settings.  
LOCal  The instrument is set to front panel control (front panel keys are active). 
REMote  The instrument is set to remote interface control (front panel keys are active). 
RWLock  The front panel keys are disabled (the instrument can only be controlled via the 
remote interface
)
.  
The remote/local state can also be set by interface commands over 
the GPIB and some other I/O interfaces. When multiple remote 
programming interfaces are active, the interface with the most 
recently changed remote/local state determines the instrument’s 
remote/local state. 
The remote/local state is unaffected by *RST or any SCPI commands 
other than SYSTem:COMMunicate:RLState. At power-on however, the 
communications setting always returns to LOCal. 
SYSTem:ERRor? 
This query returns the next error number and its corresponding 
message string from the error queue. The queue is a FIFO (first-in, 
first-out) buffer that stores errors as they occur. As it is read, each 
error is removed from the queue. When all errors have been read, the 
query returns 0, NO ERROR. If more errors are accumulated than the 
queue can hold, the last error in the queue will be -350, TOO MANY 
ERRORS (see Appendix C for error codes). 
SYSTem:VERSion? 
This query returns the SCPI version number to which the instrument 
complies. The returned value is of the form YYYY.V, where YYYY 
represents the year and V is the revision number for that year. 
*IDN? 
This query requests the power supply to identify itself. It returns a 
string of four fields separated by commas. 
 
Agilent Technologies 
xxxxxA 

<A.xx.xx>,<A.xx.xx> 
Manufacturer 
Model number followed by a letter suffix 
Zero or serial number if available 
Firmware revision, power supply revision