System Commands 
78 
Series N5700 User’s Guide  
*OPT? 
This query requests the unit to identify any installed options. A 0 
indicates no options are installed. 
*RCL <NR1> 
This command restores the power supply to a state that was 
previously stored in memory locations 0 through 15 with the *SAV 
command. Note that you can only recall a state from a location that 
contains a previously-stored state. 
NOTE 
All saved instrument states are lost when the unit is turned off. 
*RST 
This command resets the power supply to a factory-defined state. 
This state is defined as follows. Note that *RST also forces an ABORt 
command. The *RST settings are as follows: 
 
CAL:STAT Off 
[SOUR:]CURR:PROT:STAT 
Off 
INIT:CONT Off 
[SOUR:]VOLT 
OUTP Off [SOUR:]VOLT:LIM 
[SOUR:]CURR 0 
[SOUR:]VOLT:TRIG  0 
[SOUR:]CURR:TRIG 0 
[SOUR:]VOLT:PROT 
MAXimum 
*SAV <NR1> 
This command stores the present state of the power supply to 
memory locations 0 through 15.  
NOTE 
All saved instrument states are lost when the unit is turned off. 
*TST? 
This query always returns a zero.